Measuring blood pressure
Pump the cuff, open the valve slowly and listen for the Korotkov sounds. Change the patient, the cuff size and the arm’s height relative to the heart.
The “Artery” view shows the artery under the cuff opening and closing and the turbulent blood; the “Graph” shows arterial and cuff pressure and the sound phases I–V.
In this lab
- S: first sound, D: sounds vanish
- Δp = ρgh: 10 cm ≈ 7.8 mmHg
- p ≈ D + (S − D)/3
- normal ≈ 120/80
What you learn
We learn to measure blood pressure by the Korotkov method, see where the sounds come from, and test how arm height and cuff size affect the result.
Laws and formulas
S: first sound, D: sounds vanish — The Korotkov method.
Δp = ρgh — Hydrostatic pressure: 10 cm ≈ 7.8 mmHg.
p ≈ D + (S − D)/3 — Mean arterial pressure.
1 mm sim. ust. = 133,3 Pa — Pressure unit.
Equipment
- Cuff — standard 22–32 cm and large 32–42 cm.
- Aneroid gauge — 0–300 mmHg in 2 mm divisions.
- Bulb and valve — deflation 2, 4 or 8 mm/s.
- Stethoscope — in the elbow crease below the cuff.
Common mistakes
- Deflating too fast: S reads low and D high.
- Holding the arm below heart level: the reading is too high.
- A narrow cuff on a large arm: the reading is too high.
In real life
- Home automatic monitors use the pressure oscillations in the cuff (the oscillometric method).
- Hypertension is the main risk factor for stroke and heart attack.