X-rays

Switch on the X-ray tube, choose the voltage, current and aluminium filter, and place a hand phantom or a stepped Al wedge on the detector.

The “Spectrum” view shows the spectrum against λ with λ_min, the Kα and Kβ lines and a cursor; the “Image” view shows the radiograph, detector signal and contrast.

In this lab

  • λ_min = hc/(eU), hc = 1240 pm·kV
  • I = I₀·e^(−μx)
  • d½ = ln2/μ
  • W: Kα 59 keV, Kβ 67 keV

What you learn

We study the X-ray tube spectrum: find Planck’s constant from the Duane–Hunt law, see the characteristic lines, and examine attenuation and image contrast.

Laws and formulas

λ_min = hc/(eU) — The Duane–Hunt law.

hc = 1240 pm·kV — A handy form.

I = I₀·e^(−μx) — Attenuation.

d½ = ln2/μ — The half-value layer.

Equipment

  • X-ray tube — a cathode and rotating tungsten anode, 40–120 kV.
  • Collimator and filter — limits the beam; Al 0–10 mm.
  • Detector — measures the transmitted beam and forms the image.

Common mistakes

  • Losing contrast by raising the voltage too far.
  • Working without a filter gives the skin an extra dose.

In real life

  • CT rebuilds a slice from hundreds of X-ray projections.
  • Mammography uses low voltage (25–30 kV) for soft-tissue contrast.